Dissolution Testers – USP 1/2/5/6
Basic
Features

station

Infinity Test

Isolated Temperature Controller

Motorized Lift Mechanism

Quality by Design

Multi-Level Security

Multi-stage Filtration

Report

Rinsing

Sample Collector Tray

Savings

Snap-Fit Shaft
LCD display of 40 characters and 4 lines for user-friendly operation. |
Multilevel security with password. |
Alphanumeric data entry of product and user details. |
Unique TimeAction™ function allowing media changeover and infinity test run, ideal for sustained and controlled release products. |
Complies with USP, IP, Ph. Eur., and JP specifications. |
Automatic stirrer height positioning for USP 1, 2, 5, 6 intrinsic test methods. |
Easy Snap-Fit™ shaft locking mechanism for positive engagement and wobble-free operation. |
Clear water bath with a metal top plate. |
Precise individual vessel centering system. |
Adjustable leveling of the instrument. |
Vibration-free smooth motorized lift movement. |
Isolated water circulating pump for precise temperature control of the water bath. |
20 programmable protocols. |
16 programmable sampling intervals. |
Easy to empty and clean water bath with bottom drainage. |
Designed to minimize routine validation. |
Report for online validation of test, monograph, and error log during the test. |
Programmable wake-up for temperature control. |
Programmable sample volume from 1 ml to 20 ml. |
Audio-visual indicators for system status. |
Power failure detection facility. |
Split rinse to avoid carryover. |
Replenishment. |
Individual vessel temperature monitoring. |
Specifications
Model | Inspire-14 with DC |
Speed range | Variable speed from 20 to 300 RPM |
Speed accuracy | ±1 RPM |
Temperature range | 20° C to 40° C (Extended range 10°C to 40°C & 20°C to 55°C optional) |
Temperature accuracy | ±0.1°C |
Display | 40 x 4 character backlighted LCD |
No. of protocols | Maximum 16 of 1 min to 23 hrs, 59 min |
No. of samples | Serial port |
Sample interval | 80 column Dot Matrix printer |
Printer interface | RS 485/RS 232 |
Printer supported | 150ml, 250ml, 500mL, 1L |
Interface port | One piece moulded clear bath |
Vessel capacity | Heater : 1kW, SS 316 with thermal protection |
Water bath | Sensor : RTD |
Temperature controller | 220/230 VAC, 50/60 Hz,1.3 kW/6 Amp |
Electrical power | 110 VAC, 50/60 Hz,1.3 kW/10 Amp |
Dimensions (W x L x H) | 521 x 1672 x 663 mm |
Weight | 212 kg |
Ordering Information
Product | ||
Product code | Description | Quantity |
01210204 | Tablet Dissolution Tester Model Inspire-14 with Dc-230V | 1 |
01110204 | Tablet Dissolution Tester Model Inspire-14 with Dc-110V | 1 |
Standard Accessories | ||
Part number | Description | Quantity |
0104A00075 | EDT laser marked basket - 40 mesh (Set of 6) | 2 |
0104A00066 | EDT Basket Rod | 12 |
0104A00068 | EDT Basket Holder with clip (Set of 6) | 2 |
0104A00592 | Dissolution Inspire Detachable Paddle Blade(Set of 6) | 2 |
0104A00610 | Stirrer for replenishment | 2 |
0104A00242 | EDT merlon vessel - 1000ml clear (N) | 14 |
0104BV4155 | EDT V-S External Probe Without Trim Pot | 1 |
0104BV4114 | EDT- Lx V-4 Temperature Controller ETC-15 | 1 |