Melting Point
Features

User Friendly Operation

Free Drop of the Cylinder

Security

Quality by design

Results

Simultaneous Rotating and Tapping Motion

LAN


Date and Time
| Available with Stand-alone & PC through operation |
| 7" TFT touch screen display |
| 21 CFR Part 11 compliance |
| 1000 user methods in all for Melting point |
| Video recording of last sample is stored in stand-alone model |
| PC through model is necessary for storing of every sample |
| Date & Time stamped GLP Compliant 1000 Result reports with Parameters |
| Melting point & Melting range (auto/manual) of sample in terms of °C with sample result report |
| USB interface for printer, keyboard and mouse |
| LAN, WAN interface for computer |
| Perform analysis of a different unknown sample at a time in a pilot run of 50°C - 350°C |
| Automatic Detection with Video Recording |
| Video recording of melting process Every Sample analysis report comes with video analysis for observation. |
| Aluminum block with protective coating |
| A graph of intensity vs temperature is also available to graphically analyze the melting of the sample |
Specifications
| Model | EzMelt-A with EzIntegrity |
| Temperature Range | (Ambient +5)°C to 400°C |
| Temperature Accuracy | ±0.3°C (Ambient to 200°C) & ±0.5°C (200°C to 400°C) |
| Heating Time | 10 min. (50°C to 300°C) |
| Cooling Time | 0 min. (300°C to 50°C) |
| Number of Samples | 4 Samples |
| Control Module | PID Controller |
| Temperature Resolution | 0.1°C |
| Ramp Rate | 0.1°C to 20°C |
| Reproducibility | ±0.2°C |
| Mode of Detection | Automatic/Manual |
| Heater | 120W |
| Electrical Power | 220/230 VAC, 50/60 Hz, 600 W/2.6 Amp 110 VAC, 50/60 Hz, 600 W/5.5 Amp |